RFP #24374 - Biochemical Oxygen Demand Analyzer for LCPW

Addendum

Addendum #1

Q: What is the manufacturer and model number of the current BOD in use?

A: The manufacturer of our current BOD Analyzer in use is Skalar and the model number is SP-10.

Addendum #2

Q: Are one or two DO probes to be fitted on the automated BOD system?

A: One DO probe is all that is necessary, additional probes are helpful but not required.

Q: How many days are required for the onsite installation and training?

A: There is no set time requirement for installation and training. Until everything is complete, I suppose.

Q: Are BOD and cBOD required to be in the same run, or are they separated?

A: I would like BOD and CBOD to be run in the same run, that is how our current setup is. I would consider BOD and CBOD being in different runs, if necessary.

Q: What is the laboratory using for their seed source?

A: We are using InterLab for our seed source.