Q: "What will be the measurement ranges for the TOC Analyzer and TNb Module?"
A: Measurement range for TOC Analyzer can be 5-500 mg/L, measurement range for TN Module can be 1-50 mg/L
Q: "Will an Electrochemical Detector be considered for the TNb analysis?"
A: Sure, an Electrochemical Detector can be considered.
Addendum #2
Q: "Will 21.CFR Part 11 Compliant software and testing be required (Eg. IQ/OQ)? If so, will this need to be performed by the vendor?
a. The “ability to meet or exceed requirements for EPA testing methods.” can mean levels of Carbon/Nitrogen for analysis in typical EPA methods, but the EPA almost always requires methods using data-secure, audit-trail compliant software and validation testing, which is not mentioned in the document."
A: Yes.
Q: "Will a Universal Power Supply (UPS) be required to offer stable, clean power to critical instrument components in case of outages or surges to protect instrument longevity?"
A: No.
Q: "Will magnetic stirring of samples be required to maintain suspension in sample vials?"
A: No.
4. "Will the lab provide its own PC and Monitor?"
A: Yes, if necessary.